A novel device (SD-101) with high accuracy for screening sleep apnoea-hypopnoea syndrome |
言語 |
|
|
言語 |
eng |
DOI |
|
|
|
関連識別子 |
https://doi.org/10.1111/j.1440-1843.2009.01627.x |
|
|
関連名称 |
10.1111/j.1440-1843.2009.01627.x |
キーワード |
|
|
主題 |
polysomnography, portable device, pressure sensor array, screening, sleep apnoea syndrome |
資源タイプ |
|
|
資源 |
http://purl.org/coar/resource_type/c_6501 |
|
タイプ |
journal article |
著者 |
Agatsuma, Toshihiko
Fujimoto, Keisaku
Komatsu, Yoshimichi
Urushihata, Kazuhisa
Honda, Takayuki
Tsukahara, Teruomi
Nomiyama, Tetsuo
|
信州大学研究者総覧へのリンク |
|
|
氏名 |
Fujimoto, Keisaku |
|
URL |
http://soar-rd.shinshu-u.ac.jp/profile/ja.yFfaHFkh.html |
信州大学研究者総覧へのリンク |
|
|
氏名 |
Urushihata, Kazuhisa |
|
URL |
http://soar-rd.shinshu-u.ac.jp/profile/ja.jakhHVkh.html |
信州大学研究者総覧へのリンク |
|
|
氏名 |
Honda, Takayuki |
|
URL |
http://soar-rd.shinshu-u.ac.jp/profile/ja.WeDUHFkh.html |
信州大学研究者総覧へのリンク |
|
|
氏名 |
Tsukahara, Teruomi |
|
URL |
http://soar-rd.shinshu-u.ac.jp/profile/ja.yaLNbakF.html |
信州大学研究者総覧へのリンク |
|
|
氏名 |
Nomiyama, Tetsuo |
|
URL |
http://soar-rd.shinshu-u.ac.jp/profile/ja.jNACuakh.html |
出版者 |
|
|
出版者 |
WILEY-BLACKWELL PUBLISHING, INC |
引用 |
|
|
内容記述 |
RESPIROLOGY. 14(8):1143-1150 (2009) |
書誌情報 |
RESPIROLOGY
巻 14,
号 8,
p. 1143-1150,
発行日 2009-11
|
抄録 |
|
|
内容記述 |
Background and objective: The SD-101 is a non-restrictive, sheet-like medical device with an array of pressure sensors, to detect sleep-disordered breathing by sensing gravitational alterations in the body corresponding to respiratory movements. This study evaluated the accuracy of the SD-101 for screening sleep apnoea-hypopnoea syndrome (SAHS) by comparison with polysomnography. Methods: Nocturnal polysomnography and SD-101 monitoring were conducted simultaneously and compared in 201 patients with suspected SAHS (suspected SAHS group) and 165 male employees of a transport company (screening group). Results: Polysomnography revealed an AHI of <5, 5 <= AHI < 15, 15 <= AHI < 30, 30 <= AHI < 60 and AHI >= 60 events/h in 39, 35, 38, 68 and 21 subjects in the suspected SAHS group and 103, 34, 12, 12 and four subjects in the screening group, respectively. Central SAHS and obstructive SAHS were subsequently diagnosed in 11 (5.5%) and 135 (67.2%) of subjects in the suspected SAHS group and five (3.0%) and 39 (23.6%) of subjects in the screening group, respectively. Significant correlations were apparent between AHI and the respiratory disturbance index (RDI) measured with the SD-101 in both the suspected SAHS group (r = 0.88) and screening group (r = 0.92). Receiver operating characteristic curve analysis revealed 89.5% sensitivity and 85.8% specificity in identifying SAHS, using an RDI of 14.0 events/h. Conclusions: These findings suggest that the SD-101 is a useful device for screening SAHS. |
資源タイプ(コンテンツの種類) |
|
|
内容記述 |
Article |
ISSN |
|
|
収録物識別子タイプ |
PISSN |
|
収録物識別子 |
1323-7799 |
書誌レコードID |
|
|
収録物識別子タイプ |
NCID |
|
収録物識別子 |
AA11097269 |
PubMed |
|
|
|
関連識別子 |
https://pubmed.ncbi.nlm.nih.gov/19818056 |
|
|
関連名称 |
19818056 |
出版タイプ |
|
|
出版タイプ |
AM |
|
出版タイプResource |
http://purl.org/coar/version/c_ab4af688f83e57aa |
WoS |
|
|
URL |
http://gateway.isiknowledge.com/gateway/Gateway.cgi?&GWVersion=2&SrcAuth=ShinshuUniv&SrcApp=ShinshuUniv&DestLinkType=FullRecord&DestApp=WOS&KeyUT=000271064100011 |